Abstract: Ferrite chokes are often
used to reduce currents due to ESD and to provide
series impedance to suppress common mode currents that can cause EMC
emissions problems. Data is presented to show how an ESD current is
affected by single pass and multiple turn ferrite chokes. The data is
extended
to predict how the chokes affect EMC emissions. Recommendations are
inferred from the data as to when to use single pass or multiple turn
windings on ferrite cores.
Discussion: When ferrite cores, sometimes called ferrite
chokes, are used to insert an impedance in series with a conductor, the
question of using a single pass through the ferrite core or using
multiple turns often comes up. Multiple turns ideally would result in
increased impedance, a desired outcome. However, inter-turn capacitance
and other high frequency effects may cause a reduction of the series
impedance and other
problems. Data was taken using an ESD waveform to analyze performance
of
single and multiple turn ferrite chokes. An ESD waveform is ideal for
this purpose since it contains both high and low frequency components
as well as significant current that might cause core saturation in some
cases.
Figure 1 shows the test setup for evaluating multiple, six in this case, turns on a
ferrite core compared to a single pass through two ferrite cores. A
Fischer F-65 current probe is used for the current measurements. With
its one Ohm transfer impedance (one Volt output into 50 Ohms for one Ampere of current
through the probe) and one Mhz to one GHz flat frequency response, this
probe is ideal for measuring ESD currents.
The output of the ESD simulator shown in Figure 1 is being applied to the
wire passing once through two ferrite cores of the type commonly used
in EMC work. These cores have an impedance that looks something like a 150 Ohm
resistor in parallel with an inductor having an R/L corner frequency of a
few tens of MHz with the 150 Ohm resistor. A multiple turn ferrite is shown pushed out of the way on the left.
Figure 2 shows the ESD simulator applying the ESD pulse to the wire
with the six turn ferrite. The ESD was also applied to the wire on the
right with no ferrite cores to generate a baseline for the ESD
waveform. In all cases, a screw on the back panel of the oscilloscope
chassis was used as the "ground." The ESD simulator was used
without the normal IEC contact discharge tip that limits the initial
rise to about 1 nanosecond. Used in this way as shown in the Figures, with
just a straight, pointed metal tip, the ESD current has a risetime of
about 350 picoseconds. I like to use the simulator this way to get a
wider bandwidth current pulse. A generator setting of 800 Volts was
used for the test.
Figure 2. Test Setup Using a Ferrite Choke with Six Turn Winding
Figure 3 shows the current through the wire with no ferrite core, on the
right in Figure 2. The wave shape is similar to the one described in
IEC 61000-4-2 except for a second peak and other lumps on the waveform
as well as the faster risetime. The extra structure is due to the shape
of the oscilloscope chassis and its wired connections.
Notice the
waveform is an average of 16 triggers. This was done to minimize EMI
pickup because of the close range of the ESD to the scope. The EMI
manifests itself as the small noise just before the rising edge. The
averaging function minimizes its amplitude without affecting the
display of the ESD current which is very repeatable.
Figure 3. Current Through Solid Wire
(Vertical scale = 500 mA/division)
The current through two ferrites as shown in Figure 1 appears in
Figure 4. The fast initial peak is reduced to about half of its
original value, but the lower frequency part of the waveform, the
"hump" after the peak, is only slightly reduced although somewhat
smoothed compared to Figure 3. As expected, the ferrite cores used
provide more effect on the high frequency content of the waveform than
the low frequency content.
The fast parts of the waveform are usually associated with equipment
upset because the larger di/dt in those parts of the waveform generates greater voltages that can corrupt
signals. The lower frequency parts of the waveform contain most of the
energy and are often responsible for physical damage to system
components although the high peak current of the initial peak can
sometimes cause burnout as well. So it appears that the two ferrites of
Figure 1 may help with equipment upset but not so much with damage to a
system.
Figure 4. Current Through Two Single Turn Ferrite Chokes
(Vertical scale = 500 mA/division)
Figure 5 shows the current through the six turn ferrite core. Note
there is significant reduction of the low frequency components of
the current implying a reduction of the possibility of equipment
damage. However, the initial peak value of 1.5 Amperes is no better
than for the two ferrite,
single turn case of Figure 4. Note also that the peak-to-peak amplitude
of the fast spike in Figure 5 is about one third larger than for the
two ferrite single
pass case, two Amperes in Figure 5 instead of one and a half Amperes in
Figure 4. So, the possibility of equipment upset is
likely not much improved over the two ferrite case. Additional turns
would likely cause an increase of the initial peak because of increased
inter-turn capacitance.
There are other problems starting to surface in Figure 5 though. Notice
there is ringing on the waveform indicating a resonance at about 350
MHz. This may cause problems in some applications. The resonance
may also be partially responsible for the negative dip below ground,
current reversal, in the waveform. The implication of the current
reversal is that
equipment upset that may have only occurred on one polarity of an ESD
event may
now, in some cases, occur on both polarities.
Figure 5. Current Through a Ferrite Choke with Six Turn Winding
(Vertical scale = 500 mA/division)
Think about the effect of putting one or two single turn ferrites
in series with the six turn ferrite. Do you think the performance would
be good at both high and low frequencies? The answer to this and
generating the waveform to prove it is left as an exercise for the
reader.
One can easily extend these results to continuous signals. High
frequency, about 50 to 100 MHz and higher, currents are likely reduced
more by single turn ferrite chokes whereas low frequency signals,
below 10 MHz, are definitely reduced more by multiple turn ferrite chokes.
The exact frequencies depend on the ferrite material used and circuit
impedances where the ferrites are applied.
Summary:
The data presented shows that single turn ferrite chokes can be
effective at reducing the high frequency parts of an ESD current,
likely improving the system upset response to ESD. Multiple turn ferrite
chokes are more effective on the low frequency parts of the ESD current
and may help avoid system damage. These results are easily extended to
continuous signals.
tutorial on this subject, covering background as well as more technical details, is available at:
.
If you like the information in this article and others on this website, much more information is available in my courses.
to see a listing of upcoming courses on design, measurement, and
troubleshooting of chips, circuits, and systems.