Doug's picture
High Frequency Measurements Web Page
Douglas C. Smith

 Address:  P. O. Box 1457, Los Gatos, CA 95031
 TEL:      800-323-3956/408-356-4186
 FAX:      408-358-3799
 Mobile:   408-858-4528
 URL:      www.dsmith.org
 Email:    doug@dsmith.org



Technical Tidbit - March 2009
Using Current Probes to Inject Pulses for Troubleshooting - Part 2
(Current Probes, More Useful Than You Think - Part 3)


test setup with multi-turn coil on current probe

Figure 1.
Test Setup For Measuring Pulse Induced by Current Probe into Multi-turn Coil

Abstract: Current probes can be used to inductively couple voltage pulses into cables for troubleshooting without needing a direct connection to the circuit. This concept is extended to using multiple turns through the current probe to increase the injected voltage.

Discussion: In the  December 2007 Technical Tidbit, current probes were shown to be a convenient method of injecting series voltage pulses in circuits for troubleshooting. However, the amount of injected voltage may not be enough, 50 to 100 Volts was demonstrated in the December article, to cause the desired response.  Figure 1 shows the test setup used to measure the injected voltage when a wire was wrapped through the current probe for multiple turns. This will increase the injected voltage at the expense of additional series source impedance in the circuit to be stressed. Figure 2 shows a close-up of the wire wrapped on the current probe with the wire passing through the current probe four times. The induced voltage was measured using an Agilent 1163a 10:1 resistor divider probe with a 500 Ohm input impedance.

multi-turn coil around current probe

Figure 2. Close-up of Wire Wrapped on Current Probe
Fischer Custom Communications TG-EFT high voltage pulse generator was used to drive an F-33-1 current probe for this article. As shown in Figure 3, its output was set to just ~60 Volts to keep the scope from being overdriven.

tg-eft setting at ~60 Volts

Figure 3. Setting the TG-EFT at 60 Volts for the Measurements

Figure 4 shows the induced voltage peaking at about 28 Volts for a 60 Volt drive from the TG-EFT. This means that with a drive of 400 Volts from the TG-EFT, the maximum I have used with the F-33-1 current probe, injected voltages of around 200 Volts are possible. Additional turns could be added on the current probe to yield higher induced voltages. At some point, parasitic capacitance in the coil, its inductance, or physical size compared to the frequency components in the pulse will become a problem. I recommend checking the injected pulse with an oscilloscope to make sure you are getting the pulse that is desired.

Note in Figure 4 the fast rising edge and long tail which is the shape of the output pulse of the TG-EFT. The dip below zero Volts is an artifact of the current probe being AC coupled. The area under the curve must be the same above zero Volts as it is below zero Volts averaged over time.

scope plot multi turn voltage

Figure 4. Measured Pulse using Four Turns of Wire Around Current Probe
(Vertical scale = 10 Volts/div, Horizontal scale = 50 ns/div)

Figure 5 shows the same waveform expanded out from 50 ns/div to 2 ns/div to show a rise of about four nanoseconds to the peak from the baseline. The peak of ~28 Volts is more clearly seen in this view.

scope plot of multiturn coil

Figure 5. Measured Pulse Using Four Turns of Wire Around Current Probe - Expanded
(Vertical scale = 10 Volts/div, Horizontal scale = 2 ns/div)

Figure 6 shows a measurement using just one turn through the F-33-1 current probe, in this case the ground lead of the scope probe. This is similar to one of the measurements described in the December 2007 Technical Tidbit. A close-up of the current probe and scope probe appears in Figure 7.

single turn through currnent probe

Figure 6. Test Setup For Measuring Pulse Induced by Current Probe into a Single Turn


single turn through current probe

Figure 7. Close-up of Current and 1163 Resistive Divider Scope Probe

Figure 8 shows the induced voltage into the 1163A probe, for a 60 Volt drive from the TG-EFT, peaking at about 7.7 Volts. Figure 9 expands the time scale from 50 ns/div to 2 ns/div where the peak at 7.7 volts can be seen more clearly. The rise from the baseline to peak is about three nanoseconds. This is faster than the rise of four nanoseconds of the four turn case because of lowered series inductance and less capacitance to the current probe body resulting in less filtering for the single turn case. These added effects of the four turn coil are responsible for both the longer risetime and flatter peak in Figure5 versus the peak in Figure 9 for the single turn case.

The ratio of the induced voltage between four turns and one turn is 28/7.7 = ~3.6 a little lower than the turns ratio of 4:1 because the input impedance of the probe represents more of a load on the four turn coil as well as the filtering effects described in the last paragraph.


Figure 8. Measured Pulse using One Turn Around Current Probe
(Vertical scale = 2 Volts/div, Horizontal scale = 50 ns/div)


scope plot for single turn through current probe

Figure 9. Measured Pulse Using One Turn Around Current Probe - Expanded
(Vertical scale = 2 Volts/div, Horizontal scale = 2 ns/div)

The F-33-1 current probe contains just a magnetic core surrounded by a coil and handles the pulses from the TG-EFT well. Some current probes, such as the Fischer F-65 and F-33-2 probes, contain resistive networks to modify the current probe frequency and amplitude response. Probes such as these with internal resistive networks should not be used to inject pulses as the required high voltage pulses applied to the current probe may damage the resistive network.

Summary: Current probes can be used to induce series voltage impulses in cables. The induced voltage can be increased at the expense of increased risetime by increasing the number of turns of the cable through the current probe. Four turns were shown to work well for this purpose. Current probes with resistive networks in them such as the F-65 and F-33-2 should not be used in this way to avoid damaging the networks in the probes.

Additional articles on this website related to this topic are:
  1. Current Probes, More Useful Than You Think (~170K)
    (1998 IEEE EMC Symposium paper)
  2. June 1999: The Dual Current Probe Problem
  3. September 1999: Measuring Voltages Using Current Probes
  4. August 2001, Differential Measurement of Cable EMI Currents
  5. October 2007, Using Noise Injection for Troubleshooting Circuits
  6. December 2007, Using Current Probes to Inject Pulses for Troubleshooting
    (Current Probes, More Useful Than You Think - Part 2)
Equipment used in this Technical Tidbit:
  1. The scope used in this Technical Tidbit is an Agilent DSO5054A, a great little scope that is easy to use and boots in 9 seconds flat!
  2. The Agilent 1163a probe is a 10:1 resistive divider probe with over 1 GHz bandwidth.
  3. The Fischer Custom Communications F-33-1 current probe used in this article is popular with EMC engineers for measuring common mode EMI currents and is widely available in many test labs.
  4. Fischer Custom Communications TG-EFT high voltage pulse generator
If you like the information in this article and others on this website, much more information is available in my courses. Click here to see a listing of upcoming courses on design, measurement, and troubleshooting of chips, circuits, and systems. Click here to see upcoming seminars in Newport Beach, CA.

Click here for a description of my latest seminar titled (now also available online as a WebEx seminar):

EMC Lab Techniques for Designers
(How to find EMC problems and have some confidence your system will pass EMC testing while it is still in your lab).


D-104 mike
Check out my podcast containing mp3 format short tutorials, tech news and more! Just click on the microphone to see the listing of free audio programs. Content is added every week  on technical topics so check back frequently.
Top of page
Home

Questions or suggestions? Contact me at doug@dsmith.org
Copyright © 2009 Douglas C. Smith