Discussion: Figure 1 shows the
test setup used to generate the data. The setup is composed of two copper-clad boards connected by a wire about
four inches (~10 cm) long, an
Agilent N9320B spectrum analyzer, and two
Fischer F-33-1 current probes.
The two boards are separated about one mm or so by a plastic bag. One
of the current probes was connected to the tracking generator output of
the spectrum analyzer and the other was connected to the input of the
spectrum analyzer and used to monitor the current flowing between the
two copper-clad boards in the wire. The tracking generator was set to
its highest output to maximize the measured signal with respect to
measurement noise.
Also shown in Figure 1 are two 22 Ohm
resistors to be used for damping the resonance, one at a time. One of
the resistors can be seen passing through the
current probes using about 4 inches, 10 cm, of wire for the
connection between the resistor and the two boards. The other resistor can be seen at the far right side
of Figure 1. Figure
2 shows the details of the copper-clad boards, current probes, and
resistors more clearly. The resistor on the right can be placed
anywhere between the two boards since the size of the two boards are
much smaller than a wavelength at the frequencies investigated.
Figure 2. Close-up of Test Setup
Figure 3 shows the resulting plot on the spectrum
analyzer when no resistors were connected between the copper-clad
boards. There is a clear peak of tens of dB in the measured current at
about 34 MHz denoting the resonance caused by the capacitance between
the two boards and the inductance of the wire connecting them that passes
though the current probes.
Figure 3. Current Measured Between Two Copper-clad Boards at 1 mm Spacing With No Damping Resistors Used
(Vertical scale = 10 dB/div, Horizontal scale = 30 MHz/div)
In Figure 4, the result of connecting the 22 Ohm resistor on
the right side of Figure 2 between the copper-clad boards is shown. The
resonant peak of current is almost completely eliminated!
Figure 4. Current Measured Between Two Copper-clad Boards at 1 mm Spacing With 22 Ohm Damping Resistor Used
(Vertical scale = 10 dB/div, Horizontal scale = 30 MHz/div)
The result of separating the boards by about an inch, ~2.5 cm,
using a paperback book as the spacer and with no damping resistors connected
is shown in Figure 5. The increased board spacing has decreased the
capacitance between the boards and thus raised the resonant frequency
to about 75 MHz. When the 22 Ohm resistor on the right side of Figure 2
was again added, the result is shown in Figure 6. The 75 MHz resonance is
damped to the point of not being visible in Figure 6.
Figure 5. Current Measured Between Two Copper-clad Boards at 2.5 cm Spacing With No Damping Resistors Used
(Vertical scale = 10 dB/div, Horizontal scale = 30 MHz/div)
Figure 6. Current Measured Between Two Copper-clad Boards at 2.5 cm Spacing With 22 Ohm Damping Resistor Used
(Vertical scale = 10 dB/div, Horizontal scale = 30 MHz/div)
In
Figure 7, the test setup has returned to a board spacing of about 1 mm,
but the 22 Ohm damping resistor passes through the current probes. Note
that, compared to Figure 4, there is some damping, about 10 dB, but not
as much as shown in Figure 4, about 20 dB. This is due to the longer
length of the wire needed to connect the resistor through the current
probes. The added inductance of the longer wire is reducing the
effectiveness of the 22 Ohm resistor, pointing to the need to keep the
lead lengths of the resistor as short as possible to minimize
inductance.
Figure 7. Current Measured Between Two Copper-clad Boards at 1 mm Spacing With a 22 Ohm Damping Resistor Passing Through the Current Probes
(Vertical scale = 10 dB/div, Horizontal scale = 30 MHz/div)
The
value of 22 Ohms used for the resistor caused significant damping of
the resonance between the two copper-clad boards. This value will vary
for different dimensions of the boards and their spacing as well as
with the inductance of the connection between the boards. A value near
optimum can be calculated to be equal to the capacitive reactance of
the two boards at resonance (using the parallel plate capacitor
formula). A good on-line calculator for parallel plate capacitance can
be
found here.
The inductive reactance of the connection between the boards could also
be used since at resonance it is equal to the capacitive reactance, but
the inductance of the connection is more difficult to calculate than
the capacitance between the boards.